Operating Theatre Light

Operating Theatre Light

Brand Name:KENYUE Medical Product Number:KY-PLM700/500 Dimensions: Head Diameter:700 AND 500MM,WITH LED LIGHT:48PCS. Model Name:750*480*930mm/850*520*950mm Warranty Time:5 years conforms to CE, iso 13485 and ISO9001.
Send Inquiry
Description
Technical Parameters

Operating Theatre Light use LED cold light sourcewhich can be used for all kinds of surgeries. This operatinglight has the endoscopic surgery mode function.
.Head Diameter:PLM-700 :700mm ,PLM-500:500mm
LEDS:PLM-700:80 ,PLM-500:48
.llumination:40~160klx

Operating Theatre Light USING MANUAL:

1. Expected use

This product is used in various surgical lighting in hospital operating room.It can also be use in medical lighting situations in emergency room.

Conditions of use

2.1 Normal working conditions

Temperature:10℃~30℃;

Relative humidity:≤70%;

Atmospheric pressure range:860hPa~1060hPa;

This equipment can not be worked in flammable and explosive gas environment.

The type of device security: I type equipment;

Working system: continuous operation;

Liquid injection protection level: IPXO.

2.2 Power parameters

Power supply voltage:AC220(±10%)V;

Frequency:50Hz;

Input power:300VA;

PLM-500 main lamp power:88W.

2.3 suitable mounting height:3200~3300 mm

Low-storey may bring inconvenience to users. please refer to the product diagram. in special circumstances, you should contact with the manufacturer when purchasing.

Technical specifications

Type

PLM-500 Main lamp

PLM-500 Main lamp

Diameter of Light Head (mm)

500

500

Quantity of LEDS

48

48

Illumination (10 Steps Adjustable)

40~160klx

40~160klx

Color rendering index

Ra>95

Ra>95

Color temperature (5 Steps Adjustable)

3500-5000K

3500-5000K

Depth of illumination (mm)

1000

1000

Spot diameter

120-250mm

120-250mm

LED Life

>50,000 Hours

>50,000 Hours

The model of LED

OSRAM GWCS8PM1.CM

OSRAM GWCS8PM1.CM

UPS

Available

Available

Control Panel

LCD

LCD

 

Simplified diagram

FIXING

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Installation

The equipment is fixedly mounted on the roof. the floor must bear the weight of the device,but also bear the force moment because of partial load and the operation. In order to ensure the safety ,the floor should have enough bearing capacity and strength.

Users can make construction according to the following pictures.

Floor installation program

FIXING 02

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ll fasteners of the installed components must be reliable and fastening to prevent the harm due to loose. when mounting, the bottom of mounting base must be adjusted to level condition, otherwise it would cause the drift of lamp. ground wire(color :yellow or green) must be grounded reliably ,in order to ensure the use of safety.

With installed problems, please contact with the manufacturer before construction.

 

 

Hot Tags: operating theatre light, China, suppliers, manufacturers, factory, wholesale