Operating Theatre Light use LED cold light sourcewhich can be used for all kinds of surgeries. This operatinglight has the endoscopic surgery mode function.
.Head Diameter:PLM-700 :700mm ,PLM-500:500mm
LEDS:PLM-700:80 ,PLM-500:48
.llumination:40~160klx
Operating Theatre Light USING MANUAL:
1. Expected use
This product is used in various surgical lighting in hospital operating room.It can also be use in medical lighting situations in emergency room.
Conditions of use
2.1 Normal working conditions
Temperature:10℃~30℃;
Relative humidity:≤70%;
Atmospheric pressure range:860hPa~1060hPa;
This equipment can not be worked in flammable and explosive gas environment.
The type of device security: I type equipment;
Working system: continuous operation;
Liquid injection protection level: IPXO.
2.2 Power parameters
Power supply voltage:AC220(±10%)V;
Frequency:50Hz;
Input power:300VA;
PLM-500 main lamp power:88W.
2.3 suitable mounting height:3200~3300 mm
Low-storey may bring inconvenience to users. please refer to the product diagram. in special circumstances, you should contact with the manufacturer when purchasing.
Technical specifications
|
Type |
PLM-500 Main lamp |
PLM-500 Main lamp |
|
Diameter of Light Head (mm) |
500 |
500 |
|
Quantity of LEDS |
48 |
48 |
|
Illumination (10 Steps Adjustable) |
40~160klx |
40~160klx |
|
Color rendering index |
Ra>95 |
Ra>95 |
|
Color temperature (5 Steps Adjustable) |
3500-5000K |
3500-5000K |
|
Depth of illumination (mm) |
1000 |
1000 |
|
Spot diameter |
120-250mm |
120-250mm |
|
LED Life |
>50,000 Hours |
>50,000 Hours |
|
The model of LED |
OSRAM GWCS8PM1.CM |
OSRAM GWCS8PM1.CM |
|
UPS |
Available |
Available |
|
Control Panel |
LCD |
LCD |
Simplified diagram

Installation
The equipment is fixedly mounted on the roof. the floor must bear the weight of the device,but also bear the force moment because of partial load and the operation. In order to ensure the safety ,the floor should have enough bearing capacity and strength.
Users can make construction according to the following pictures.
Floor installation program

ll fasteners of the installed components must be reliable and fastening to prevent the harm due to loose. when mounting, the bottom of mounting base must be adjusted to level condition, otherwise it would cause the drift of lamp. ground wire(color :yellow or green) must be grounded reliably ,in order to ensure the use of safety.
With installed problems, please contact with the manufacturer before construction.